Q:How do I upgrade the software?
Software upgrades are available from Scanivalve Corp. A detailed procedure may be found in the DSA3200 Series Software Manual, which is also available for download from this website. For more information on updating software for these modules, contact Scanivalve Corp, Product Service Department.
Q:How do I change the IP Address?
The IP Address in a DSA3200 series module may be changed by a user. The IP address is stored as a variable in the operating system. This variable can be changed only by using the RS232 connection. A procedure for this may be found in the DSA3200 Series Software Manual.
Q:What are the power requirements?
DSA3217 and DSA3207 modules require 28 ± 4 Vdc at 10 VA.
DSA3016 modules are powered by the DSA3016 enclosure
DSA3218 modules with an optional heater require 28 ±4 Vdc at 35 VA.
Q:I have lost my backup coefficients. What can I do?
Scanivalve Corp keeps a record of the coefficients for each module generated in the most recent calibration performed at Scanivalve Corp. These coefficients can be obtained at no charge by contacting Scanivalve Corp, Product Support Dept.
Q:How often should I calibrate the module?
DSA modules should be re-calibrated at a six month interval for the best accuracy. Scanivalve Corp provides a utility program for customers who can perform their own calibrations. This program is called PressCal and may be downloaded from this website at no charge.
Q:I cannot communicate with the Unit on the Ethernet connection?
Verify that the Ethernet cable is the correct cable for the connection. A “straight wired” cable must be used if the Host computer is connected to the DSA module through a hub. For direct connections from a Host to a DSA module, a “cross wired” cable must be used.
The Host computer must have an IP address specified to be in the same class as the DSA module. For example: if the DSA module has an IP address of 191.30.5.100, the Host computer must have an IP address of 191.30.80.sn.